Thermo 15C HCl Analyzer
The Thermo 15C HCl analyzer is ideal for monitoring at lower ranges (0-50ppm). It is highly specific to HCl and contains its own pump. It is recognized for its long term zero and span stability. This item is also available for rent from CleanAir Europe.
Overview
Range:
0 -10 up to 0 - 5,000ppm
Power:
105 - 125 v AC, 60 Hz / 220 - 240 v AC, 50 Hz
Shipping Weight:
50 lbs.
Helpful Hints
Helpful Hints:
The Thermo 15CHL should be used by experienced stack testers only. When moisture is removed from the gas sample, there is a tendency to lose HCl due to its high solubility in water. Clean Air has had success using a permeation dryer with a heated pump in correct applications with this instrument. Dilution is also an option. A traditional gas conditioning system cannot be used. To help prevent cell corrosion, please purge with nitrogen after each use.
Download Detailed Specifications
Detailed Specifications: